30-500MA 、HF、DENTAL X RAY 、C-ARM
500mA Micro-computer Control X-ray Equipment JY-A17
Detailed Product Description
500mA Micro-computer Control X-ray Equipment:
1 Micro-compute control x-ray radiography assembles: optional 500mA.
2 Total 540 storage combination controlled by element such as kV mA S is applied radiography for the following filed three body position, fat middle and thin body type, automatic detect and display cod of mistake in system, send warning signal, detect accuracy and maintain.
3 X-ray tube capacity protection, anode starting-up protection and radiograph time-limit protection device semi-conduct controlled, fluoroscopy computed time device, x-ray tube over-hot protectorate provided, ensure x-ray tube run safely long time, improve accuracy of kV mA S.
4 Radiography table can move lengthways and crossover, can loaded all kinds of radiography dark box, use conveniently and it is beautiful.
5 Applied in all kinds of hospital for radiography such as general bodybeam filtrationand chest.
Department
|
Technology parameters
|
||||||||
High voltage generator control table and high voltage transform assemble
|
Power
|
Voltage
|
380V±10%
|
||||||
Capacity
|
50Hz±1Hz
|
||||||||
Frequency
|
≥32KVA
|
||||||||
Inner resistance
|
≤0.3Ω
|
||||||||
Tube voltage
|
45-125kV step 1kV
|
||||||||
Tube currency
|
30-500mA 8 shift
|
||||||||
Exposure time
|
0.01-5.0S 25 shift
|
||||||||
Max rated capacity
|
mA
|
600
|
500
|
400
|
500
|
400
|
|||
kV
|
71
|
84
|
106
|
84
|
106
|
||||
X-ray assembly
|
Type of x-ray
|
XD52-30.50/125
|
|||||||
Focus
|
1×1.2×2 double focal
|
||||||||
Rated rotate speed of anode
|
2800r/min
|
||||||||
Manually beam limited device
|
Model
|
Manually continuous adjusted field of
|
|||||||
Field view
|
View, luminosity≥100LX,1minlimit-time device
|
||||||||
CDG
Multi function radiography table
|
Moving table
|
Lengthways
|
±600mm
|
Electromagnetism locking
|
|||||
Across
|
±120mm
|
||||||||
Mobile beam filtration grid
|
Grid density
|
28L/cm
|
|||||||
Grid ratio
|
81
|
||||||||
Focusing distance
|
100cm
|
||||||||
Scope of moving
|
≥600mm
|
||||||||
Radiography item
|
5×7,8×10,10×12,11×14,
12×15,14×14,14×17
|
||||||||
Distance from surface to film
|
75mm
|
||||||||
Height of bed
|
700mm
|
||||||||
Moving pillar
|
±180°,per 90°orientation, moving lengthways ≥1500mm
|
||||||||
Bulb tube
|
Rotate the cross arm±180,ascend and descend 1100mm
|
||||||||
|
|
|
|
|
|
|
|
|
|
Option
LSJ pillar radiography frame
|
The scope of ascend and descend
|
1450mm
|
|
The minimum distance from focal spot
|
380mm
|
||
Radiography item
|
×10 10×12 11×1412×15
14×14 14×17
|
||
Fixed beam filtration role
|
Grid density
|
40L/cm
|
|
Grid radio
|
10:1
|
||
Focal distance
|
180cm
|