Radiology Room |
Ultrasound Room |
Surgery Room |
Laboratory Room |
Comprehensive Room |
Pediatrics Room |
Dental Room |
Medical operation instruments |
Hospital Furniture |
Medical supplies |
30-500MA 、HF、DENTAL X RAY 、C-ARM
3.3KW Mobile High Frequency X-ray Machine JY-20
Detailed Description
Detailed introduction to 3.3KW Mobile High Frequency X-ray Machine:
MAIN TECHNICAL PARAMETER
|
||
x-ray generator
|
HF4003
|
|
high-frequency generator
|
40kHz
|
|
ripple
|
≤3%
|
|
total filtration
|
above 2.7 mmAl
|
|
target angle, focuse
|
12°,1.5 mm
|
|
power
|
3.3 kW
|
|
thermal capacity
|
665kHU
|
|
Max. mA and mAs
|
70mA/320mAs
|
|
max voltage
|
110 kV
|
|
power 230V(±10%) 50/60Hz
|
18A
|
|
weight
|
122 kg
|
|
Mode of operation and functions
|
||
part of Control
|
microprocessor control table installed in the mobile support
|
|
User Interface
|
The ergonomic control panel and backlight LCD
|
|
Photo manually
|
2 technical parameter(kV,mAs)
|
|
Anatomical hotography
|
LP and HP all 20 sites to choose
|
|
X-ray control
|
Second gear manual switch
|
|
safty
|
-mA min and mA max Safety chain
-the largest loaded chain of x-ray tube and the maximum exposure time security protection
-X-ray tube components temperature thermal protection and capacity computing
-kV min ,kV max,Δ kV max,I max safeguard
-Start rotating anode protection
-Microprocessor self-test
|