20KW High Frequency X-ray Machine JY-A27
1.Maximum output power
Intermittent mode: 100kV ,200mA,20kW。
2.Adjustment range of tube current、tube voltage
tube current: Radiography 16mA~200mA。
tube voltage: Radiography 40kV~125 kV。
3.Exposure Time:photography0.0025s~6.4s step adjustable
4 Nominal Frequency≥30kHz
5.The condition of power
Power Voltage:380V±10%
Power supply capacity: 120% the power of the equipment
Internal resistance:≤0.5Ω。
6.High frequency and high voltage x-ray generator(including microcomputer control console)
Nominal frequency of high voltage x-ray generator≥30kHz。
Internal resistance: less than 0.5Ω.
Nominal Output power:20kW。
Output Voltage: Radiography: 40 kV~125kV, step 1kV;
Output Current: Radiography: 16mA~200mA,
Exposure Time: 0.0025s~6.4s。
Display:digital pipe display photography voltage\ Photography current\ photography time\ mAS of photography
X-ray tube
Max tube voltage:125kV Max tube current:400mA。
Small focus:1.0mm ;Big focus:2.0mm。
Speed of anode rotating:2800r/min
Anode heat capacity:140KHU
7.Collimator
Collimator with fixed position lamp.
Power supply: Working voltage: 24V
8.Radiography table
Integration of the table with stand column without skyrail design.
Length*width*height of table surface:2000mm×760mm×690mm. Deviation:±10mm
Distance from table surface to film:50mm
Horizontal movement range of table≥220mm
Vertical movement range of table≥900mm
has locking function, electromagnetic braking, Braking force: ≥98N
9.X-ray source component pillar:
With scale that can indicate the height of the x-ray tube unit
Distance of vertical movement:770mm±10mm;+90°~-90°, deviation<±2°
Moving crosswise (left & right):+180°~-180°, deviation<±2°
Moving lengthways (along ground railway) ≥1300mm
10.Bucky: Fixed beam filtration grid density≥40 lines/cm.
grid ratio 10:1,。Focus Distance:110cm
11. Horizontal movement of bucky ≥500mm
Electromagnetic braking
Braking force ≥98N,
Button control
12. Collimator open and closedown should be flexible, reliable,can adjust to any Radiation in field rectangle size.
13. Stand chest stand
Fixed beam filtration grid density≥40 lines/cm.
Ray-filter grid:grid ratio 10:1, Focus Distance:110cm
Cassette size:5″×7″ 8″×10″ 10″×12″ 11″×14″ 14″×17″