Radiology Room |
Ultrasound Room |
Surgery Room |
Laboratory Room |
Comprehensive Room |
Pediatrics Room |
Dental Room |
Medical operation instruments |
Hospital Furniture |
Medical supplies |
200MA x ray machine
200mA Conventional Radiogrphy X-ray Unit JY-A10
Main Characteristics of Medical X Ray Machine
● Single-table / Single-tube
● Rotary anode X-ray tube unit and tangential annular tube head
● Single-phase full wave rectification high-voltage generator
● Power voltage (V) radiograph kilovolt (KV). Step-less control electric mechanism
● Filament stabilizer for X-ray tube and space charge complementor
● Radiograph volume KV/mA and S. grading and interlock protection
● Adopt digital circuit timer. Grading according to R10 Priority coefficient. (for more accurately circuit of silicon-controlled rectifier of heavy power)
● Radiographic table can move in length and breadth.
● The radiographic table, column and vibrating grid are in a whole without ceiling and floor rails bottom
Main Technical Indexes of Medical X Ray Machine
● Single-table / Single-tube
● Rotary anode X-ray tube unit and tangential annular tube head
● Single-phase full wave rectification high-voltage generator
● Power voltage (V) radiograph kilovolt (KV). Step-less control electric mechanism
● Filament stabilizer for X-ray tube and space charge complementor
● Radiograph volume KV/mA and S. grading and interlock protection
● Adopt digital circuit timer. Grading according to R10 Priority coefficient. (for more accurately circuit of silicon-controlled rectifier of heavy power)
● Radiographic table can move in length and breadth.
● The radiographic table, column and vibrating grid are in a whole without ceiling and floor rails bottom
Main Technical Indexes of Medical X Ray Machine
Item
|
Contents
|
Index
|
Demand of Power Supply
|
Capacity
|
Two-phase 30KVA
|
Voltage Frequency
|
380V/220V ±10% 50HZ ±0.5HZ
|
|
Internal Resistance
|
0.75Ω (380V) 0.25Ω(220V)
|
|
Power Supply
|
Adjustment Range
|
380±10% at 380V,
220V±10% at 220V, being adjustable continuously |
Radiograph
|
Voltage
|
50—125KV being adjustable continuously
|
Current
|
Small Focus 50mA 100mA
Big Focus 50mA 100mA 200mA |
|
Time
|
0.04~6.35s 23grades together, according to R10 Coefficient
|
|
HT Generator
|
Capacity
|
20KVA(Instantaneous)
|
Max DC Output Voltage
|
125KV
|
|
Max DC Output Currency
|
200mA
|
|
X-ray Tube
|
Model
|
XD51. 20—40 / 125
|
Focus
|
Small Focus 1×1 mm Big Focus 2×2 mm
|
|
Radiographic Table
|
Length×Width×Height of Bed Surface
|
2000mm×665mm×700mm
|
Moving range of bed surface
|
640mm in Length, 200mm in breadth
|
|
Ray-filter of bed
|
Travel in Length≥500mm
|
|
Grid Density N=28
|
||
Convergence Distance : f=100cm
|
||
Grid Ratio : R=8
|
||
Column of X-ray Tube Unit
|
Moving Longitudinally along Radiographic Table
|
1000mm or 1800mm
|
Distance to ground as moving up and down (along column)
|
650~1800mm
|
|
Rotary around the center of cross arm
|
±180°
|
|
Rotary around the shaft line of X-ray Annular Tubes
|
-10° ~ -60°~ +120°
|
|
Max Size of Cassette
|
|
356mm×432mm (14°×17°)
|