200MA x ray machine
Detailed Product Description
200mA medical x-ray equipment:
Features of Medical X-ray Machine:
● Our high quality medical x-ray machine has double diagnostic table and double X-ray tube with rotated anode.
● KV, mA and a interlocking protection and radiograph time-limited protection devices are provided.
● Filament for stabilizing voltage, space charge compensation and anode starting-up protection device are provided.
● Voltage of power supply and kV with digital display can be adjusted constantly and respectively.
● This medical x-ray machine becomes convenient to install without above-ground railway and be operated easily.
● Electromagnetism are provided for pillar lengthways, crosswise, and sleeve moving up and down.
Technology Parameters
Description |
Technology Parameters |
|
Power Supply |
Voltage |
A.C.380V / 220V (±10%) |
Frequency |
50Hz ± 1Hz |
|
Capacity |
20KVA |
|
Power Supply Resistance |
A.C.380V≤1.0Ω; A.C.220V≤0.3Ω |
|
Maximum tube voltage |
120KV;125KV |
|
Maximum tube currency |
200mA |
|
Fluoroscopy |
Tube voltage |
40~110KV |
Tube currency |
0.5~5mA |
|
Radiograph |
Tube voltage |
40~120KV; 40~125KV |
Tube currency |
Small Focus:50mA;100mA; Big Focus:100;150;200mA |
|
Spot film |
100mA (Small Focus) |
|
Time |
Divided into 23 grades from 0.04~6.3s |
|
High Voltage Generator |
Capacity |
20KVA |
X-ray Tube |
Model |
XD 51—20, 40 / 125 |
Focus |
Small Focus 1×1mm2 |
|
Rotated Speed of anode |
2800r / min |
|
Table for radiograph |
Length×width×height |
2000×760×680(mm) |
Movable scope of the bed |
Lengthways: ±450mm; Crossways: ±110 |
|
Pillar moves lengthways |
1000mm |
|
Distance from surface to film |
≤60mm |
|
Bulb tube moves up and down |
1100mm~1760mm |
|
Radiography item |
10"×12" 11"×14" 12"×15" |
|
Diagnostic Table |
Scope of rotation |
90°~ 0°~-5° |
Stomach & Intestine Radiograph Device |
Spot film device specification(Stomach & Intestine Radiograph) |
127×178mm( 5"× 7") |
203×254mm( 8"×10") |
||
254×305mm(10"×12") |
||
Moving lengthways (along ground railway) |
1500mm |
|
Moving crosswise (left & right) |
740mm |
|
Moving up and down |
700mm ~ 1900mm |
|
Way of locking |
Electromagnetism locking |
|
Beam Limited Device |
Manually single automatic multi-optional |