Radiology Room |
Ultrasound Room |
Surgery Room |
Laboratory Room |
Comprehensive Room |
Pediatrics Room |
Dental Room |
Medical operation instruments |
Hospital Furniture |
Medical supplies |
10-100MA x ray machine
100mA Medical X-Ray Machine JY-A2
100mA medical x-ray machine
USE
Fluoroscopy, Digestive track spot film radiography, General radiography, Filter-ray device radiography
Unit Configuration
Control, HT generator unit, Ht cable, X-ray tube unit, Diagnostic table, Column
MAIN CHARACTERIDTICS
1. Single table and single tube
2. Design according to the electric mechanism of middle-size machine ,The function is reasonable
3. The high-tension system is of split structure. Equipped with afullwave rectification high-voltage generator and independent X-ray tube unit, with large output power and reliable and stable operation
4. Supply voltage (V), roentgenoscopy, radiography (kv) and stepless adjust able
5. Equipped with the manostat for filament of X-ray tube and the spacecharge compensator as to ensure the stable and accurate output
6. Equipped with the capacity limitation and portection circuit and KV mA S interlock limitation and protection
7. Adopt the digial circuit timer, grade according to R10 priority coefficient and can control time exactly,
8. Diagnostic table is of electric back rotation and is equipped with vibrating filter-ray device
9. The spot film device of luminescent serene can hold the film jecket of various sizes and can make the photography of full-film and two sectional films separateiy.
10.The column post hast has the bottom track only, which is easy while installation
MIN TECHNICAL INDEXES
Fluoroscopy, Digestive track spot film radiography, General radiography, Filter-ray device radiography
Unit Configuration
Control, HT generator unit, Ht cable, X-ray tube unit, Diagnostic table, Column
MAIN CHARACTERIDTICS
1. Single table and single tube
2. Design according to the electric mechanism of middle-size machine ,The function is reasonable
3. The high-tension system is of split structure. Equipped with afullwave rectification high-voltage generator and independent X-ray tube unit, with large output power and reliable and stable operation
4. Supply voltage (V), roentgenoscopy, radiography (kv) and stepless adjust able
5. Equipped with the manostat for filament of X-ray tube and the spacecharge compensator as to ensure the stable and accurate output
6. Equipped with the capacity limitation and portection circuit and KV mA S interlock limitation and protection
7. Adopt the digial circuit timer, grade according to R10 priority coefficient and can control time exactly,
8. Diagnostic table is of electric back rotation and is equipped with vibrating filter-ray device
9. The spot film device of luminescent serene can hold the film jecket of various sizes and can make the photography of full-film and two sectional films separateiy.
10.The column post hast has the bottom track only, which is easy while installation
MIN TECHNICAL INDEXES
ITEM
|
CONTENTS
|
INDEX
|
||
Upright post(column)
|
Moving longitudinally(along the bottom track)
|
1500mm
|
||
Moving latitudinal
|
360mm
|
|||
Moving up and down
|
from focus to ground
|
|||
Supply power
|
Voltage
|
220V/380(single phase )
|
||
Frequency
|
50Hz±0.5Hz
|
|||
Capacity
|
10KVA
|
|||
Intemal resistance
|
380v:2,220V:0.6
|
|||
Max DC output voltage
|
50-90KV
|
|||
Max DC output current
|
100mA
|
|||
Fluoroscopy
|
Voltage
|
50-90KV
|
||
current
|
≤5mA
|
|||
Radiograph
|
Voltage
|
50-90KV
|
||
current
|
Small focus 15mA , big focus30mA 60mA 100mA
|
|||
spot film
|
100mA(fixed)
|
|||
time
|
(0.04-6.3s,24grades together)
|
|||
High-voltage generator
|
Capacity
|
7KVA
|
||
X-Ray tube
|
Model
|
XD4-2.9/100
|
||
Focus
|
Small focus 1.8mm big focus4.3mm
|
|||
Diagnostic table
|
rotary angle
|
+90°~0~-5°
|
||
specification of spot film (stomach and intestines photography)
|
128×179mm(5”×7”)
|
|||
205×254mm(8”×10”)
|
||||
254×305mm(10”×12”)
|
||||
MAX VOLUME OF RADIOGRAPH
|
||||
Focus of X-Ray tube
|
Radiograph mA
|
Max Radiograph KV
|
Max exposure time allow
|
|
Small focus
|
15mA
|
90KV
|
6.3S
|
|
Large focus
|
30mA
|
90KV
|
6.3S
|
|
60mA
|
90KV
|
6.3S
|
||
100mA
|
85KV
|
2.5S
|