Reason analysis of tube damage
When we repair medical-x-ray machine, the tube damage cause the faults. But there are several sides reasons of tube damage:
First, Open the filament or half open
Fault phenomena is that add the high pressure, but no tube current or current is not stable. For x-ray machine tube, only cathode filament heat to the ability of launching electronic and add positive high pressure between cathode and anode; Then there must be current through between cathode and anode. If there is high pressure but no tube current and filament power supply normally, tube filament open in most conditions. Then you can measure filament resistance to confirm directly. But you must be attention to that sometimes filament is locate in half open condition, sometimes it connects and sometimes it disconnects. Heating causes open when it works and it connects when it doesn’t have power; It will cause misunderstanding.
Second, Rotating anode doesn’t turn
When happens this fault, it displays rotating anode fault. It is fault which the tube itself causes, there are two reasons causing this fault: One is rotating anode stator coil damage; The other is anode rotor rotating doesn’t work or large clearance after bearing damage. When it charges, Magnetic field distribution is imbalance.
Third, Vacuum tube core decreases or tube core glass shell damage
This fault is that overload alarm. There are two reasons causing tube core vacuum decreasing: One is that tube itself or machine control parts causing metal target surface damage or vaporization, it leads to vacuum degree decrease in tube; The other is that tube doesn’t use for a long time and there is gas entering into tube core and tube core damage make oil go into tube.
Fourth, High voltage spark of tube set inside
The fault happens, alarm display overload; The slight is that there is pa, pa high voltage spark voice from tube. Although this machine can work, the tube has been damaged. You must stop using.
Fifth, The tube leakage
Seals damage causes the tube leakage.